NYAD SERIES 400
CARBON DIOXIDE ANALYZER
FlashPaper
Series 400 Photo

 FEATURES

CO2 Sensor technology

Many gases, including CO2, absorb energy in the infrared band.  This absorption is selective and occurs at specific frequencies corresponding to the resonant frequencies of bond vibrations within the molecule.  Measuring a characteristic absorption enables the gas to be detected and the strength of the absorption gives a measure of the gas concentration.
TheCO sensor optics consists of an infrared source, a sampling cell, an infrared filter, and a detector.  In addition, the sensor includes electronics to drive the source and to process the signals from the detector.


The infrared source is the hot filament of a tungsten lamp emitting a broad band of radiation, which then passes through the gas in the sample cell then through a filter before reaching the detector.  The filter selects out a very narrow bank of frequencies corresponding to a characteristic absorption band for CO2. The level of energy reaching the detector falls as the concentration of CO2 increases.

SPECIFICATIONS

Sensor
Range: 0-2,000 PPM
Accuracy: At +25 °C (+77°F)
Operating Temperature: -5 to +45°C
Power Consumption: Less than 3 watts
Non-linearity: <1% FS
Response Time: 63% in 1 min.

Enclosure
NEMA-4: 9.5"W x 6.25"H x 3.5" D
Panel: 10"W x 5.25"H x 6"D
Rack: 19"W x 5.25"H x 6"D

Electronics
Display: 4 Digit LCD, 0.5" High
Alarm: SPDT, 1A @ 120 VAC
Output: 0 to 10 VDC, or 4-20 mA
Sample Flow: 100 to 250 cc/min
Power: 120 VAC 50/60 Hz.,1W