NYAD
SERIES 600
MOISTURE/CARBON
MONOXIDE ANALYZER
FEATURES
- Independent analyzers
for each measurement
- Microprocessor
based electronics
- Wide
dynamic range, Moisture: -130°C to +20°C Dew/Frostpoint
- Selectable units: (°C, °F, or ppm,
Moisture)
- Direct
reading Carbon Monoxide concentration: 0-2000 ppm
- Auto calibration (CO)
- Adjustable alarm contacts
- Analog output (Voltage or Current)
- Two 4 digit LCD displays
- O.E.M. packaging
TWO
CRITICAL MEASUREMENTS IN ONE PACKAGE
The NYAD Series 600 Analyzers
were designed to accurately and reliably measure both moisture and
Carbon Monoxide concentrations in breathing air or medical air supplies
as well as provide alarms when safe levels are exceeded.
For the
first time in one package, the 600 series offers two independent
analyzers with separate displays, outputs and alarms.
The A.C.T.
(Advanced Capacitive Transducer) moisture sensor uses state-of-the-art
technology to produce a rugged, fast responding, trace moisture sensor
that is also designed to lower your costs by permitting sensor
replacement without having to discard old probe bodies.
The
A.C.T. technology has a high specificity to moisture as well as
dramatic improvements in its performance compared to older, competitive
versions. The A.C.T. sensor is mounted on a probe containing digitizing
electronics and calibration data in electronic memory - this means all
moisture probes are completely interchangeable.
The Carbon
Monoxide Analyzer features an advanced electrochemical sensor whose
digitized output is sent to a microprocessor based analyzer which
provides a digital display of Carbon Monoxide concentration, up to two
software adjustable alarm contacts with visual and audible alarms, and
a voltage or current analog output that has user adjustable zero and
span. In addition, an automatic calibration feature allows the user to
easily perform periodic calibrations using certified span gases.
SPECIFICATIONS
Sensor
Moisture Sensor
Type:
NYAD A.C.T., Series 50
Range: -130°C to +20°C Dew/Frostpoint
Accuracy:
± 2°C
Repeatability: ± 0.5°C
Operating Temp: -130°C to +70°C
Pressure
Rating: 7000 psig
CO Sensor
Type:
NYAD COS-1
Range: 0-2000 ppm
Resolution: 0.5% of Range
Accuracy:
± 1%
Sample Flow: 100 to 200 cc/min.
Enclosure
Rack: 19"W x 5.25"H x 11.5"D
Panel:
10"W x 5.25"H x 11.5"D
NEMA-4: 11"W x 15"H x 5"D
Electronics
Display: 4 Digit LCD, 0.5" High
Alarm:
SPDT, 1A @ 120 VAC
Output: 0 to 10 VDC, or 4-20 mA
Power: 120 VAC
50/60 Hz., 1W Max.

200 Mason Circle, Suite A, Concord, CA 94520
Ph: (925) 270-3971, Fax: (925) 270-3973 www.nyad.com